首頁
1
商品介紹
2
CP-6001 6,000A FLEXIBLE CURRENT PROBE3
https://www.gtcl.url.tw/ 格創科技有限公司
首頁 商品介紹 POWER ANALYZER CP-6001 6,000A FLEXIBLE CURRENT PROBE
1066

CP-6001 6,000A FLEXIBLE CURRENT PROBE

60 A/600 A/6000 A, 3 ACV at full scale, probe length 900 mm

CP-6001 an cooperate any kinds multimeters ( ACmV meters )

6,000A FLEXIBLE CURRENT PROBE

Model : CP-6001                                                            ISO-9001, CE, IEC1010

543cec350fad854ede6eb584991f64ad.png

FEATURES 

* Range : 60 A/600 A/6000 A. 

* Output sensitivity : 50 mV/A, 5 mV/A, 0.5 mV/A.

* Output voltage : 3 ACV ( 60 A, 600 A, 6000 A ).

* Flexible probe length : 900 mm ( 35.4 inch ).

* Optional current probe for DW-6093, DW-6095.

* When change the new probe, the calibration procedures are not necessary to be executed.

* Signal output use the banana plugs, durable and safety.

* Power supply is from DC 1.5 V,  AA ( UM-3 ) battery  x 2 PCs or  from the optional  AC to DC 9V adapter. or from the meter (DW-6093, DW-6095 ) via Power interface cable, PWCB-06.

SPECIFICATIONS 

Measurement ranges

60 A/600 A/6000 A, 3 ranges, switch selectable

   1st range value : 60 A

   2nd range value : 600 A

   3rd range value : 6000 A

Output sensitivity

50 mV/A, 5 mV/A, 0.5 mV/A,

AC coupled.

Accuracy

 ± 1% of range value

   * Accuracy is specified that the   measured conductor's position is on the center of " Flexible current probe ".

   * 50/60 Hz.

   * 23 ± 5

Frequency range

45 to 400 Hz.

   * Accuracy tested under 50/60 Hz.

Linearity

 ± 0.2 % of reading value from 10 %  to 100% of range value.

Phase error

± 1.5°

   * 50/60 Hz.

Conductor position sensitivity  

± 2 % of range value.

   * Recommend the measured conductor  position should  > 25 mm ( 1" )  away from the " Probe locker ".

External field influence

increase ± 1.5 % of range value max..      

   * Recommend the distance between different "  Flexible current probe " should > 200 mm ( 8" ).

Minimum load

100 K ohm  for specified accuracy.

Enclosure material

UL 94-V0 rated. 

Safety 

IEC 1010, Categories III 600V.

Power Supply

1 DC 1.5 V,  AA ( UM-3 ) battery x 2 PCs

2 AC to DC 9V power adapter ( optional ).

3 Power supply from the meter ( DW-6093, DW-6095 ) via Power interface cable, PWCB-06 ( included ).

   * Power supply voltage from DW-6093, DW-6095 is DC 8V± 1V.

OperatingTemperature

 0  to 50 ( 32 to 122 ).    

Operating Humidity

 Less than 80% RH.

Dimension

Main circuit unit : 

    105 x 56.5 x 33 mm.

    ( 4.1 x 2.2 x 1.3 inch ).

Flexible probe : 

     900 mm x 14.5 mm DIA.  

     35.4 inch x 0.57 inch DIA.

Cable length :

     3.3 meters ( 10.8 feet ).

DC plug size :

     5.5 mm x 2.5 mm DIA.

Power Consumption

Approx DC 9 mA.

Weight

523 g ( 1.2 LB ).

Accessory Included

Operating manual ...........................1 PC.

Power interface cable, PWCB-06.......1 PC.

Optional Accessory

AC to DC 9V power adapter.

FRONT PANEL DESCRIPTIONS 

c913e4603270bf9dbaa09c8db069988a.png
1367018