FC-2700 FREQUENCY COUNTER
2.7 GHz, Bench type, with TCXO
FREQUENCY COUNTER
Model : FC-2700 ISO-9001, CE, IEC1010
FEATURES
TCXO (temperature compensated crystal oscillator) time base, high stability & accuracy. |
High sensitivity for the VHF & UHF frequency measurement, useful for the CB amateur. |
Wide measuring range up to 2.7 GHz. |
Used the exclusive Microprocessor IC offered the intelligent function: Frequency, Period, Multi resolution, Data hold, Relative measurement, Data record ( Max., Min., Average reading ). |
8 digits, 18.3 mm large LCD . |
0.1 Hz resolution for 10 MHz. |
LCD display for low power consumption & clear read-out even in bright ambient light condition. |
Power supply from battery or AC to DC 9V adapter. |
RS 232 PC serial interface. |
GENERAL SPECIFICATIONS
Display | 18.3 mm ( 0.72" ) LCD (Liquid Crystal Display), 8 digits. | |
Measurement | Frequency, Data hold, Relative, Memory ( max., min., average ), Period. | |
Range | 2.7 GHz | 100 MHz to 2700 MHz |
100 MHz | 10 MHz to 100 MHz. | |
10 MHz | 10 Hz to 10 MHz | |
Period | 10 Hz to 10 MHz | |
Resolution, Sample Time | Ref. the following "Table for Resolution & Sample Time". | |
Sensitivity
( Sensitivity Sw. set to high position ) | 10 MHz &Period | 30 mV rms. ( 10 Hz to 10 MHz ) |
Typical : 1 5 m V r m s . ( 10 Hz to 9 MHz ) | ||
100 MHz | 50 mV rms. ( 10 MHz to 100 MHz ) | |
Typical : 2 5 m V r m s . ( 30 MHz to 100 MHz ) | ||
2.7 GHz | 50 mV rms. ( 100 MHz to 2.5 GHz | |
Typical : 3 5 m V r m s . ( 300 MHz to 2.4 GHz ) | ||
Max. functional signal input ( Sensitivity Sw. set to normal position ) | 10 MHz & period | 15 V rms |
100 MHz |
| |
2.7 GHz |
( 400 MHz to 2.7 GHz ) | |
Over-input ( Max. signal will not hurt the circuit ) | 10 MHz & Period range : | |
2.7 GHz & 100 MHz range: Max. 4 V rms. | ||
Time Base Stability vs. Temp | 1.5 PPM ( 10 to 30 ). |
Frequency Accuracy | ± ( 2 PPM + 1 d )
| |
Time Base circuit | 16.777216 MHz, TCXO (temperature compensated crystal oscillator).
| |
Input Connector | 10 MHz & Period range : | |
100 MHz range : N coaxial connector. | ||
2700 MHz : N coaxial connector. | ||
Case | Durable & strong ABS-plastic housing with handle. | |
Operating Temp. | 0 to 50 ℃ ( 32 to 122 ). | |
Operating Humidity | Less than 80%. | |
Power Supply | 6 x 1.5 V AA ( UM-3 ) battery or AC to DC 9V adapter.
| |
Power Consumption | 2700 MHz & 100 MHz range : Approx. DC 105 mA.
| |
10 MHz & Period range Approx. DC 45 mA.
| ||
AC Adapter Power Input | Optional, 9V DC , 300 to 500 mA rating, central positive for socket. | |
Dimension | 280 x 210 x 90 mm ( 11.0 x 8.3 x 3.5 inch ). | |
Weight | 1200 g/0.27 LB (including battery). | |
Standard Accessories | Instruction Manual .............. 1 PC. | |
OptionalAccessories | PB-21 | Direct probe with BNC connector & alligator clip pairs, available for 10 MHz range |
BB-22 | Direct probe with double BNC 10 MHz range | |
NN-23 | Direct probe with double N coaxial connector, available for 100 ,2700 MHz range | |
NB-24 | N coaxial connector to BNC | |
UPCB- 02 | Isolated RS232 cable. |
TABLE FOR RESOLUTION & SAMPLE TIME
Range | Gate Time Select | Resolution | Sampling Time |
10 MHz | FAST | 10 Hz | 0.5 SEC |
SLOW | 1 Hz | 1.25 SEC | |
SLOW (select 1) | 0.2Hz |
| |
SLOW (select 2) | 0.1 Hz | 11 SEC | |
100 MHz | FAST | 100 Hz | 0.75 SEC |
SLOW | 10 Hz | 6 SEC | |
SLOW (select 1) | 20 Hz | 5 SEC | |
SLOW (select 2) | 50Hz |
| |
2700 MHz ( 2.7 GHz ) | FAST | 1000 Hz | 0.5 SEC |
SLOW | 100 Hz | 2.75 SEC | |
SLOW (select 1) | 200 Hz | 1.5 SEC | |
SLOW (select 2) | 500 Hz | 0.75 SEC |