首頁
1
商品介紹
2
GCH-2018 Carbon dioxide, Humidity, Temp., NDIR, CO 2 METER3
https://www.gtcl.url.tw/ 格創科技有限公司
首頁 商品介紹 CO2 METER GCH-2018 Carbon dioxide, Humidity, Temp., NDIR, CO 2 METER
1147

GCH-2018 Carbon dioxide, Humidity, Temp., NDIR, CO 2 METER

Carbon dioxide, Humidity, Temp., NDIR, 0 to 4,000 ppm, RS232 

CO 2 METER

Model : GCH-2018                          ISO-9001, CE, IEC1010
6839a4d64852db41d55d8b7f4eb4cb6f.png

FEATURES

* NDIR method principal for CO2 ( Carbon dioxide ) measurement, available for long term operation.

* High repeatability and high accuracy.

* Two probes, one is for CO2/Temp. measurement, the other probe is for Humidity/Temp./Dew point measurement.

* Separate probe, easy operation and convenient for remote measurement.

* CO2 function with alarm setting.

* Humidity measurement with fast response time.

* Large S-TN LCD, high contrast, easy readout.

* Data hold function for freezing the desired value on display.

* Records Maximum and Minimum readings with Recall. 

* RS232/USB computer interface.

* Microprocessor circuit assures maximum possible accuracy, provides special functions and features.

* Heavy duty & compact housing with hard carrying case, designed for easy carry out & operation.

* Auto shut off is available to save battery life.

* Power supply from batteries or DC 9V adapter in.

 

GENERAL SPECIFICATIONS

Circuit

Custom one-chip of microprocessor LSI circuit.

Display

LCD size : 52 mm x 38 mm

dual function LCD display.

Measurement

CO2 ( Carbon dioxide ), Temp.

Humidity, Dew point, Temp. 

Unit

CO2

ppm

Humidity 

% RH 

Dew point   

, 

Temp.

, 

Response Time

 

 

 

 

 

 

 

 

 

CO2 :

< 2 min. typically.

   @ Reach the 63% reading value

   @ Depend the environment air

       circulation.

Humidity/Dewpoint 

10 to 30 seconds typically.

   @ Reach the 85% reading value

   @ Depend the environment air

      circulation.

CO2 altitude

compensation

setting

0 to 9,000 meters.

 

 

 

Temperature

Compensation

Automatic temp. compensation.

 

 

Adavanced

setting

 

 

 

CO2 altitude value setting 

CO2 alarrm value setting

/ setting

Auto power off enable/disable setting 

Alarm setting 

For CO2 measurement only.

Data Hold 

Freeze the display reading.

Memory Recall 

Maximum & Minimum value.

Display

Sampling Time 

Approx. 1 second.

 

Power off

 

Auto shut off saves battery life or 

manual off by push button.

Data Output 

 

 

 

 

 

RS 232/USB PC serial interface. 

* Connect the optional RS232 cable 

   UPCB-02 will get the RS232 plug.

* Connect the optional USB cable

   USB-01 will get the USB plug.

Probes no. 

 

 

 

 

Two probes :

* Probe 1 is for CO2, Temp. measurement

* Probe 2 is for Humidity , Dewpoint

  Temp. measurement.

Operating 

Temperature

0 to 50

Operating

Humidity

 

 

Main instrument : Less than 85% R.H. 

C02 probe : Less than 85% R.H. 

Humidity probe :  0 to 95 %RH. 

Power Supply 

 

DC 1.5 V battery ( UM3, AA ) x 6 PCs, 

or equivalent. 

Power Current 

 

 

 

 

 

CO2 measurement 

Approx. DC 9.6 mA for 90% period.

Approx. DC 128 mA for 10% period. 

Humidity measurement 

Approx. DC 5.6 mA. 

Weight

 

 

 

 

Main instrument : 372 g/0.82 LB. 
@ Battery is included.

CO2 probe : 158 g/0.35 LB. 

Humidity  probe : 82 g/0.18 LB. 

Dimension 

 

 

 

 

 

 

 

Main instrument : 

    173 x 68 x 42 mm 

    (7.9 x 2.7x 1.2 inch) 

CO2 Probe : 

    185 x 38 x 26 mm

Humidity Sensor Probe : 

     200 x 23 x 19 mm

Accessories

Included

 

 

Instruction manual........................1 PC 

CO2 probe....................................1 PC 

Humidity probe.............................1 PC

Hard Carrying case.......................1 PC

Optional

Accessories

 

 

ACV to DC 9V adapter 

RS232 cable, UPCB-02 

USB cable, USB-01 

Data Acquisition software, SW-U801-WIN 

 

 

 ELECTRICAL SPECIFICATIONS  (23±5 )
0e267106cf79064aa2ab525349e15774.png 

8960e88c4cc27bdc8b037b89cd0ae17c.png 

Dew Point 

   

Range

-25.3 to 48.9

Resolution

0.1 

    

Range

-13.5  to 120.1 .  

Resolution

0.1 .

Remark:    * Dew Point display value is calculated from the

                   Humidity/Temp. measurement automatically. 

                * The Dew Point accuracy is sum accuracy value of
                   Humidity & Temperature measurement.

 

1367762