HV-40 HIGH VOLTAGE PROBE
DCV - 0 to 40 KV, ACV - 0 to 28 KV rms
HIGH VOLTAGE PROBE
Model : HV-40 ISO-9001, CE, IEC1010SPECIFICATIONS | ||
Attenuate Ratio | 1 : 1000. | |
Input Impedance | Approx. 1000 M ohm. | |
Output Impedance | Around 1.1 M ohm. The input impedance of external voltmeter should be 10 mega ohm. | |
Safety | Meet CAT II 40000 V | |
Max. Working Voltage | DCV | DC 40 KV |
ACV | Peak AC 40 KV or 28 KV rms ( depend which values is larger ). | |
Accuracy
| DCV | 1 KV to 20 KV : ± 1 %. > 20 KV to 40 KV : ± 2%. |
ACV | 1 to 28 KV rms, 50/60 Hz : 5 % to -10 %. | |
Temp Coefficient | Less than 200 ppm/℃ | |
Operating Temperature | 0 ℃ to 50 ℃ ( 32 ℉ to 122 ℉ ). | |
Operating Humidity | Less than 80% RH. | |
Cable Length | 1 meter. |
OPERATION
Connect the plugs to the volts ( Hi ) & com ( Lo ) input terminals of your voltmeter ( or Multimeter ). Select the desired range of voltmeter ( Attention : Do not use auto ranging ).
Whenever possible, turn the high voltage source off before making any connections. Connect the HV probe common lead ( alligator clip ) to a good earth ground or reliable chassis ground.
SAFETY PRECAUTION & WARNING !!!
* This high voltage probe must be used by the person who are trained only. Do not work alone when working with high voltage circuits & environment.
* For your own safety, inspect the probes for cracks & frayed or broken leads before each use.
If any defects are noted, do not use the probes.
* Hands, shoes, floor & work bench must be dry. Avoid making measurements under humid, damp or other environmental conditions that might affect the safety of measurement situation.
* The ground connection must always be made before the probe tip comes into contact with the high voltage & must not be removed until after the probe tip has been removed from high voltage source.
* Do not attempt to take measurement from sources where the chassis or return lead is not ground.
* If possible, always turn the high voltage source off before connecting or disconnecting the probe.
* Before turning the high voltage on, make sure that no part of your body is in contact with the device under test.
* The probe body should be kept clean & free of any conductive contamination. Clean only the exterior probe body & cables. Use a soft cotton cloth lightly moistened with a mild solution of detergent & water. Do not allow any portion of the probe to be submerged at any time.