TC-424 THERMOMETER CALIBRATOR

THERMOMETER CALIBRATOR
source & measurement
Model : TC-424                      ISO-9001, CE, IEC1010

88d1ccba880ceaf4392ce2d1c7ba105c.png

FEATURES

* Four kinds Source output : 

  1. Pt-100 source
  2. Voltage( DC mV) source
  3. 400 ohm source
  4. Type K/J/R/E/T/S Temp. source

* Four kinds measurement inputer :

  1. Pt-100 measure
  2. Voltage( DC mV) measure
  3. 400 ohm measure
  4. Type K/J/R/E/T/S Temp. measure

* Portable instrument for calibrating process devices.

* Four kinds Source Adjustable output.

 

General Specifications

Circuit

Custom one-chip of microprocessor LSI circuit.

Display

LCD size : 52 mm x 30 mm.

LCD with white backlight ( ON/OFF ).

Measurement

1.Type k/J/R/E/T/S Temp. , 2. Voltage (DC mV). 3. Pt-100 ohm, 4. 400 ohm .

Source

Type k/J/R/E/T/S Temp.,

Voltage (DC mV : - 10.00 to + 110.00 mV),

Pt-100 ohm , 400 ohm .

Sampling Time of Display

Approx. 1 second.

Advanced setting

* Auto power OFF management.

* Set unit  or  .

Operating Temperature and Humidity

0 to 50 .

Less than 85% R.H.

Power Supply

* Alkaline or heavy duty DC 1.5 V battery ( UM3, AA ) x 6 PCs, or equivalent.

* DC 9V adapter input. ( AC/DC power adapter is optional ).

Power Current

Normal operation ( w/o current source ) :

Approx. DC 18 mA.

LCD Backlight is ON) : Approx. + DC 2 mA.

Weight

Meter : 282 g/ 0.61 LB. (with cable)

Dimension

Meter : 198 x 68 x 45 mm

Accessories Included

* Instruction manual...............................................................1 PC

* Hard carrying case( CA-06 )................................................1 PC

* ( Type K source output line ), AC-DUK02............................1 PC

* MTL-CP424..........................................................................1 PC

Optional Accessories

AC to DC 9V adapter.

TP-01,TP-02A, TP-03, TP-04, TP-05

 

Electrical Specifications ( 23.0 ± 5 )

 

Thermocouple

94438e6f5c91abcd34071e28d48568b9.png


DC Voltage

Range of source / measure

Display Resolution

Accuracy source / measure

-10.00mV to +110.00 mV

0.01 mV

± ( 0.05 % + 0.03mV )

* input impedence : 10 ^12 ohms.

* measure protection : DC 60 V , AC 24 V.

* source protection : Reference CAUTION.


Pt100

Range of source / measure

Display Resolution

Accuracy

-200.0 to 850.0 

0.1 degree

source

measure

-328.0 to 1562.0 

0.1 degree

±(0.05%+0.6)

±(0.05%+1.1)

±(0.05%+ 1)

±(0.05%+ 1.8)

* input impedence : 10 ^12 ohms.

* protection : Reference CAUTION.

Remark :
The above specification are tested under the environment RF Field Strength less than 3 V/M & frequency less than the 30 MHz only.

 

400 Ω

Range of source / measure

Display Resolution

Accuracy

source

measure

400.0 Ω

0.1 Ω

±(0.1%+ 0.4Ω)

±(0.1%+ 0.6Ω)

* input impedence : 10 ^12 ohms.

* protection : Reference CAUTION.

* FS : full scale

Remark :
The above specification are tested under the environment RF Field Strength less than 3 V/M & frequency less than the 30 MHz only.

 

4882f1e7e10eb0709d93181455f31499.pngCAUTION

* Do not apply a voltage exceeding the maximum input voltage, otherwise the input part may be damaged.
* Do not short_circuit or apply an external voltage to output terminals of the instrument or standard equipment, or else their internal circuitry may be damaged.