TC-920 THERMOMETER CALIBRATOR
THERMOMETER
CALIBRATOR
Model : TC-920 ISO-9001, CE, IEC1010
FEATURES
- 2 in 1, one meter included " type k thermometer " & " type k thermocouple calibrator ".
- Precision calibrator for type k thermometer, calibrating process devices and measuring process signals.
- Wide range : -199℃ to 1230℃ ( -325℉ to 2246℉ ).
- Microprocessor circuit assures high accuracy and provides special functions and features.
- Build in temperature linearity compensation circuit, high precision.
- ℃ & ℉ select & data hold function for the temperature measurement.
- ℃ & ℉ select by the front button for the temperature calibrator.
- Thermocouple sensor for Temp. measurement, fast response time.
- RS 232 PC serial computer interface.
- Temp. measuring input to meet any standard type K ( NiCr-NiAl ) probe.
- Fitted with standard type K input measuring socket.
- Super large LCD display, easy reading.
- Operates from 006P DC 9V battery.
- Used the durable, long-lasting components, including a strong, light weight ABS-plastic housing case.
GENERAL SPECIFICATIONS
Circuit | Custom one-chip of microprocessor LSI with thermocouple linearity correction circuit. | |
Display | 14.8 mm (0.58") digit size. Super large LCD display with annunciator. | |
Function | 2 in 1 : Type k thermometer, Type k thermocouple calibrator. | |
Range | Measure : -199.9℃ to 1230℃/ - 327℉ to 2246℉. | |
Calibrator: -199.9℃ to 1230℃/ - 327℉ to 2246℉. | ||
Polarity | ' - ' indicates negative polarity. | |
Thermometer | Sensor Type | Thermocouple K ( NiCr-NiAl ). |
Input Impedance | 10 Mega ohm. | |
Sampling | Approx. 1 second. | |
Data Output | RS 232 PC serial computer interface. | |
Over input indication | Indication of " ". | |
Operating Temperature | 0 ℃ to 50 ℃ ( 32 ℉ to 122 ℉ ). | |
Operating Humidity | Less than 80% RH. | |
Power Supply | 006P DC 9V, MN1604/PP3 battery or equivalent. Heavy duty or Alkaline type. | |
Power Current | Approx. DC 9 mA. | |
Weight | 280 g/0.62 LB ( included battery ). | |
Size | 185 x 78 x 38 mm ( 7.3 x 3.0 x 1.5 inch ). | |
Accessories Included | Instruction Manual ............................ 1 PC. Dual type k plug cable, DUK-02.......1 PC. | |
Optional Accessories | * Temperature probe, TP-01, TP-02A, TP-03, TP-04. * Carrying case, CA-03, CA-04. * Software, SW-U801-WIN. * RS232 cable, UPCB-01. * USB cable, USB-01. |
ELECTRICAL SPECIFICATIONS (23±5 ℃)
Type K thermocouple Calibrator
Calibrator Range | Resolution | Accuracy | |
℃ | -199.9 ℃ to -100.1 ℃ | 0.1 ℃ | ± (1 % + 1 ℃) |
-100.0 ℃ to 999.9 ℃ | 0.1 ℃ | ± (0.5 % + 1 ℃) | |
1000 ℃ to 1230 ℃ | 1 ℃ | ||
℉ | -327 ℉ to -200 ℉ | 1 ℉ | ± (1 % + 2 ℉) |
-199.9 ℉ to -148.1 ℉ | 0.1 ℉ | ||
-148.0 ℉ to 999.9 ℉ | 0.1 ℉ | ± (0.5 % + 2 ℉) | |
1000 ℉ to 2246 ℉ | 1 ℉ |
Thermometer, measure
Thermometer Range | Resolution | Accuracy | |
℃ | -199.9 ℃ to -100.1 ℃ | 0.1 ℃ | ± (1 % + 1 ℃) |
-100.0 ℃ to 999.9 ℃ | 0.1 ℃ | ± (0.5 % + 1 ℃) | |
1000 ℃ to 1230 ℃ | 1 ℃ | ||
℉ | -327 ℉ to -200 ℉ | 1 ℉ | ± (1 % + 2 ℉) |
-199.9 ℉ to -148.1 ℉ | 0.1 ℉ | ||
-148.0 ℉ to 999.9 ℉ | 0.1 ℉ | ± (0.5 % + 2 ℉) | |
1000 ℉ to 2246 ℉ | 1 ℉ |
The accuracy specification of thermometer applies only to the instrument itself and allowance must be made for limits of error permitted in thermocouple.